Homepage | Login | Want to open an account?
Product Search
Quick Find

Chemical oxygen demand analysis set : Thermoreactors AL 32 & AL 38

Chemical oxygen demand analysis set : Thermoreactors AL 32 & AL 38
 Cat. NoDescrtiptionPack SizePriceQuantity
WT/75014 8 PLACE REACTOR FOR 16 MM DIAMETER VIALS WITH PLASTIC COVER each £665.28
WT/75016 25 PLACE REACTOR FOR 16 MM DIAMETER VIALS WITH PLASTIC COVER each £947.23

Chemical Oxygen Demand (COD) is a critical environmental parameter and is widely used as a measure of the pollutants in discharged waste waters. COD is also used for assessing the efficiency of water treatment plants and the load placed on these plants by incoming waste streams.
COD analysis gives a rapid indication of the level of both organic and inorganic oxygen demanding pollutants. COD testing must be performed under tightly controlled conditions of temperature and time.
Reagecon offer all of the components required for rapid, confident analysis of COD. The measurement system complies with ISO 15705 & Standard Method 5220 D for chemical composition and for the requirement that digestion must take place at 150°C for 2 hours. Reagecon’s COD set comprises of:
  • The Aqualytic PC Compact COD Vario Photometer.
  • 8 place & 25 place Thermoreactors (AL32 & AL38).
  • 16mm diameter COD Reagent vials, covering 3 different measurement ranges.
  • A comprehensive range of COD standards. Thermoreactors AL 32 & AL 38 The AL 32 reactor is fitted with 8 holes for 16 mm diameter vials. For other chemical digestion operations the AL 32 reactor also includes a 22 mm hole. The AL 38 reactor is fitted with 25 holes for 16 mm diameter vials. Supplied complete with plastic protective cover.
    Technical specifications
    Model AL32 AL38
    Housing  Round metal block in epoxy covered housing,¬8 holes 16mm Ø,¬1 hole 22mm Ø Round metal block in epoxy covered housing,¬25 holes 16mm Ø,¬
    Front Robust membrane keypad, Robust membrane keypad,
    Keypad 2 keys 2 keys
    Temp. Settings 70°C / 100°C / 120°C / 150°C ± 0.3°C
    Timer 30, 60, 120 minutes and continuous operation, automatic switch-off, with alarm
    Heating 270 Watt electronically controlled, with protection against overheating 400 Watt electronically controlled, with protection against overheating
    Heat up Time 10 minutes approx. 10 minutes approx.
    Power Supply 230V / 115V / 50-60Hz 230V / 115V / 50-60Hz
    Weight 2.15kg 3.6kg
    Dimensions 214 x 133 x 110mm 275 x 155 x 95mm




  • Your Order
    Your cart is empty.


    Please Note: All our prices are subject to VAT.
    Tel England UK: +44 (0)15395 36830 ~ email: sales@eurodyne.co.uk